ClearanceBrief
Recall Z-1912-2012 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

DePuy Orthopaedics, Inc.

ULTIMA*LX STM,POL 10/12TAP/2M Packaging: The product was packaged within inner and outer PETG blister trays and sealed with coated Tyvek lids. Foam protectors were in direct contact with the stem to help immobilize the product within the package. The sterile barrier system was then packed within a carton and shrink wrapped. Intended Use: The ULTIMA LX Cemented Polished Femoral Hip Stems product

DateJun 29, 2012 (initiated May 16, 2012)
StatusTerminated
ReasonDePuy Orthopaedics, Inc. is issuing a Field Safety Notice for the LUSTER Cemented Polished Femoral Hip Stems and the ULTIMA LX Cemented Polished Femoral Hip Stems that were implanted with any metal-on-metal (MoM) articulation (including a head of CoCr and an acetabular cup where the cup is of an all metal construction (monoblock or modular)).The LUSTER Cemented Polished Femoral Hip Stems and the U
Root causeDevice Design
ActionDePuy sent an "URGENT RECALL INFORMATION-FIELD SAFETY NOTICE" letter dated May 16, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Contact the firm at 574-372-7333 for questions regarding this notice.
Quantity431 ULTIMA LX Cemented Polished Femoral Hip Stems in 78 lots
DistributionNationwide Distribution-including the states of CA, CO, IA, NC, OH, TX, WA, WI, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K972435ULTIMA AND ULTIMA LX CEMENTED FEMORAL STEM
Johnson & Johnson Professionals, Inc., Sep 11, 1997