Recall Z-1756-2012 ยท JDW Pin, Fixation, Threaded
Smith & Nephew Inc
JET-X(R) 6 MM X 60 MM SHORT HALF PIN TI-6AL-4V, REF 71056608, QTY: (1), STERILE R, Smith & Nephew, Inc., Memphis, TN 38116 Orthopedic surgical device.
| Date | Jun 6, 2012 (initiated Apr 13, 2012) |
| Status | Terminated |
| Reason | Plastic trays which encase these products have burrs on the hinged packaging which can puncture the single barrier sterile pouch, voiding sterility. |
| Root cause | Packaging |
| Action | Smith & Nephew Sales Representatives and Smith & Nephew International Distributors were notified of the problem with the affected product and the recall via email on April 13, 2012. All affected direct consignees were notified via Fed Ex on April 13, 2012. The Urgent Medical Device Recall letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to cease use and distribution of the product and quarantine them immediately for return. Customers were instructed to complete the Inventory Return Certification Form included with the letter |
| Quantity | 27 units |
| Distribution | Worldwide Distribution - USA (nationwide) and the countries of Australia, Malaysia, Germany, Japan, Portugal, United Arab Emirates, Spain, France, South Africa, Norway, Sweden, Denmark, Switzerland, Belgium, China, The Netherlands, Finland, Mexico, Italy, UK, Venezuela, Colombia, Thailand, Chile, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K994143 | SMITH & NEPHEW EXTERNAL FIXATION SYSTEM [UNILATERAL (LINEAR) & MULTILATERAL (CIRCULAR) FIXATORS AND ACCESSORIES] Smith & Nephew, Inc., Feb 18, 2000 |