ClearanceBrief
Recall Z-2031-2012 ยท JTS Stains, Microbiologic

Becton Dickinson & Co.

BD Gram Stain Kit, containing 1- 250 mL bottle of Gram Crystal Violet, and additional reagents: Gram Iodine (Stabilized), Gram Decolorizer, Gram Safranin. Gram Stain Kits and Reagents are used to stain microorganisms from cultures or specimens by the differential Gram method.

DateJul 18, 2012 (initiated Apr 12, 2012)
StatusTerminated
ReasonPotential for increase level of artifacts to appear as large, spherical and easy to distinguish or small & more difficult to distinguish from gram positive cocci.
Root causeMaterial/Component Contamination
ActionBD Diagnostic Systems sent an "URGENT PRODUCT RECALL" letter dated April 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue use and discard the affected product. An Acknowledgement Form was attached for customers to complete and return via fax to 410-316-4258. Contact the BD Customer Service Department for questions regarding this recall.
Quantity539 kits
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.