Recall Z-1706-2012 ยท NKB Thoracolumbosacral Pedicle Screw System

Stryker Spine

Stryker Xia 3 Anti-Torque Key Manual Surgical Instrument; Catalog Number 48237026. Product Usage: : Reusable instrument used to aid in the final tightening of the spinal construct. Intended Use: The Anti-Torque Key allows the Torque Wrench to align with the tightening axis and maximizing the torque needed to lock the implant assembly during final tightening of the spinal construct.

DateJun 5, 2012 (initiated Jan 12, 2012)
StatusTerminated
ReasonStryker received four reports regarding a handle which separates at the point where it is welded to the shaft on the Xia 3 Anti-Torque Key during surgery.
Root causeMixed-up of materials/components
ActionStryker Spine sent an Urgent Product Recall letter dated January 12, 2012 with an attached Customer Response forms via FEDEX to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine their inventory to identify the affected product and return using the prepaid mailing label provided. The letter asked customers to complete the Customer Response Form and fax to 201.760.8370. For questions call 201.760.8298.
Quantity95 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071373STRYKER SPINE XIA III SPINAL SYSTEM
Stryker Spine, Aug 27, 2007