ClearanceBrief
Recall Z-2092-2012 ยท MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer

Zimmer Inc.

Nextgen Complete Knee Solution Replacement Jaw for Use with Femoral and Provisional Impactor/Extractor, REF 00-5901-026-40, Non-Sterile Zimmer, Warsaw Indiana The instrument is attached to the femoral implant or provisional to aid in positioning and inserting the components.

DateJul 27, 2012 (initiated Apr 30, 2012)
StatusTerminated
ReasonThe design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design. The spring clip may fracture or fall off of the instrument during use, and a device fragment may be left in the surgical site. The risk of disassembly or fracture occurs when the instrument is typicall
Root causeComponent design/selection
ActionZimmer sent an Urgent Medical Device Recall letter dated April 30, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Requests for alternative instruments should be requested by calling (574) 372-4470. Fax Back Inventory Return Certification Form is to be sent back to Zimmer at (574) 372-4265. All recalled instruments available at the distributor are to be returned to Zimmer Product Service Department, 1777 West Center Street, Warsaw, IN 46580. The Distributor notice included instructions to notify surgeons immedia
Quantity65 units
DistributionWorldwide Distribution - USA (nationwide) and the countries of Singapore, China, Hong Kong, Canada, Australia, Japan, Europe, Germany, India, and Finland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031061NEXGEN POROUS, UNCEMENTED FEMORAL AND TIBIAL BASEPLATE COMPONENTS
Zimmer, Inc., Oct 9, 2003