ClearanceBrief
Recall Z-1692-2012 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Brilliance Big Bore - Model #728243, and - Model #728244, CT Scanning System is subject to recall/correction Product Usage: The Brilliance 6, 16, 40 64, Brilliance Big Bore, iCT, iCT SP and Ingenuity CT are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes

DateMay 30, 2012 (initiated Mar 26, 2012)
StatusTerminated
ReasonComponent defect. Following their receipt of a customer complaint, Philips Medical determined that the metal blocks which support the scissor slide rails underneath the scanning table (couch) of several of the firm's CT scanning systems were not manufactured to specification and may crack.
Root causeNonconforming Material/Component
ActionPhilips issued an 'URGENT Medical Device Correction' letter dated March 27, 2012 to all affected customers. The letter identified the affected products, problem, hazard involved, how to identify affected products, and actions to be taken by customer/user. The letter informed customers that Philips intends to dispatching service personnel to each customer site in order to correct this issue. In the meantime, the letter recommend customer to contact their Philips representative if they hear unusual noises when moving the couch vertically, or if there is unintended vertical motion of the couch.
Quantity12 system units were distributed in the U.S.
DistributionWorldwide Distribution - U.S. (Nationwide) including the states of: CA, FL,, HI, IA, IN, KY, LA, MA,, MI, MO, MS. NM, NV, OH, PA, PA, SC, TN, TX, WA, and Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012009MX8000 IDT CT SYSTEM
Marconi Medical Systems, Inc., Sep 13, 2001
K033326PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM
Philips Medical Systems, Inc., Oct 29, 2003
K033357ACQSIM-MULTISLICE-CT, MODEL 453567042271
Philips Medical Systems (Cleveland), Inc., Jan 9, 2004
K060937BRILLIANCE VOLUME
Philips Medical Systems (Cleveland), Inc., Jun 5, 2006