Recall Z-1676-2012 ยท MAX Intervertebral Fusion Device With Bone Graft, Lumbar

Alphatec Spine, Inc.

SOLUS Anterior Lumbar Interbody Fusion Spinal Spacer System, Solus Proximal/Distal Deployment Wrenches, 510(k) K102402. The Alphatec Solus Anterior Lumbar Interbody Fusion (AUF) System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DOD) at one or two contiguous levels from L2-S1 with up to Grade 1 spondyiolisthesis or retrolisthesis at the

DateMay 29, 2012 (initiated Mar 19, 2012)
StatusTerminated
ReasonAlphatec Spine has received several complaints (n+4) on the Alphatec Solus Distal Deployment Wrench for twisting or bending, which has been discovered either during or after instrumentation use. A fifth complaint received in which it was reported that the spline tip of the Alphatec Solus Proximal Deployment Wrench had sheared from the instrument upon use.
Root causeDevice Design
ActionAlphatec Spine sent an Urgent Medical Device Recall Notification dated March 30, 2012, to all affected customers. The notification letter informed customers of the removal/recall and included instructions on what to do with the recalled products. Initial letter was delivered via FedEx, and followed by telephone call with a total of 4 attempts. Customers were instructed to return all of the affected product. Upon receipt of of the Medical Device Recall Notification, customers were instructed to contact Alphatec Spine's Customer Service Department immediately for instructions on how to return
Quantity18 devices
DistributionWorldwide Distribution - USA (nationwide) and Germany
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102402SOLUS ANTERIOR LUMBAR INTERBODY FUSION (ALIF) SPINAL SPACER SYSTEM
Alphatec Spine, Inc., Mar 30, 2011