Recall Z-1701-2012 ยท CFP Radioimmunoassay, Immunoreactive Insulin

Siemens Healthcare Diagnostics

IMMULITE 2000 --- Insulin --- Catalog Number: L2KIN2 (200 test), L2KIN6 (600 tests) --- Test Code: INS, Color: Orange --- Product Usage: For in vitro diagnostic use with IMMULITE and IMMULITE 1000 Systems Analyzers --- for the quantitative measurement of insulin in serum or heparinized plasma, for the management of diabetes.

DateJun 2, 2012 (initiated Apr 30, 2012)
StatusTerminated
ReasonSiemens Healthcare Diagnostics has observed a negative bias in the Immulite Insulin assay.
Root causeComponent change control
ActionSiemens sent an Urgent Device Recall letter dated April 2012 via fax to affected U.S. customers and an Urgent Field Safety Notice to customers outside the U.S. The letter identified the affected kit lots, reason for voluntary recall, risk to health and actions to be taken by the customer/user. Customers were instructed to discard any inventory of the affected kit lots and complete and return the attached Field Correction Effectiveness Check questionnaire within thirty (30) days to request a no-charge replacement kits. Customers were requested to post this field correction by each IMMULITE s
DistributionWorldwide Distribution - US (nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022603IMMULITE 1000 AUTOMATED IMMUNOASSAY ANALYZER
Diagnostic Products Corp., Aug 21, 2002