ClearanceBrief
Recall Z-1849-2012 ยท IZL System, X-Ray, Mobile

Medtronic Navigation, Inc.

Medtronic 0-arm Imaging System with Software Version 3.1.1, Catalog Numbers: BI-700-00027-120R and BI-700-00027-120. The 0-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging.

DateJun 21, 2012 (initiated Oct 12, 2010)
StatusTerminated
ReasonSoftware Version 3.1.1 does not meet navigational accuracy.
Root causeSoftware design
ActionMedtronic Navigation contacted customers by telephone on 10/12/2010 and with "Urgent Field Safety Notice" follow up letters dated 10/14 or 10/15/2010. Sites were visited by Medtronic Service Representative to complete navigational accuracy testing and either replace or correct the unit.
Quantity3 units
DistributionNationwide Distribution including the states of CO, IN, and MN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092564O-ARM IMAGING SYSTEM
Medtronic Navigation, Inc. (Littleton), Jun 17, 2010