ClearanceBrief
Recall Z-1813-2012 ยท MKQ Processor, Cervical Cytology Slide, Automated

Tripath Imaging, Inc.

PrepStain Installation Kit (PreCoat), Catalog/Model # 490528

DateJun 15, 2012 (initiated Mar 22, 2012)
StatusTerminated
ReasonThe PrepStain Syringing Pipettes may not dispense the minimum volume of 7.5 mL of the sample due to leaking. This event resulted in an increase in complaints of leaking syringing pipettes
Root causeComponent design/selection
ActionBD Diagnostics sent a "BD PrepMate Syringing Pipette" letter dated March 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The firm encouraged customers to refer to the Operator's Manual and inspect the centrifuge tubes to verify that the correct amount of sample has been transferred. Customers experiencing this issue were instructed to discard the pipettes and contact BD Diagnostic Technical Support for a replacement. Contact the firm at 1-877-822-7771 for questions regarding this notice.
Quantity46,112,784 for all products in Recall Event
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.