Recall Z-1653-2012 ยท MUI Media, Coupling, Ultrasound
Pharmaceutical Innovations, Inc.
Other-Sonic Generic Ultrasound Transmission Gel 5 Liters Hypoallergenic Product Usage is a non-sterile gel used in ultrasound procedures to improve the transmission of the ultrasound signal from the transducer to the body.
| Date | May 24, 2012 (initiated Mar 2, 2012) |
| Status | Terminated |
| Reason | Other-Sonic GENERIC ULTRASOUND TRANSMISSION GEL is being recalled due to possible microbial contamination. An FDA Safety Communication was issued for this product on April 18, 2012. |
| Root cause | Employee error |
| Action | Phamaceutical Innovations Inc. sent an Urgent Message Notice letter dated March 02, 2012 with a followup letter dated March 9, 2012 and an Inventory Withdrawal letter dated March 23, 2012 via email and fax to all affected customers. Customers were instructed to examinne inventory, refrain from distribution and return remaining inventory for a replacement. For questions call 973-242-2900. |
| Quantity | 40,164 units in distribution worldwide - unable to obtain units in distribution Nationwide |
| Distribution | Worldwide Distribution - United States (nationwide) including the states of: FL, MA, MS, NJ, OH, TN and VA, and the countries of: Canada and Australia. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.