Recall Z-1650-2012 ยท ISH Component, External, Limb, Ankle/Foot

College Park Industries, Inc.

Celsus prosthetic foot, Product Number 103658-00. The Celsus is intended for low impact level lower limb prosthesis users weighing up to 300 lbs.

DateMay 23, 2012 (initiated Apr 5, 2012)
StatusTerminated
ReasonAn incorrect shorter part ( pyramid) was installed in the foot. A foot of this size and configuration requires a longer pyramid. Customers requiring the extra firm configuration may have a minimal risk of component failure, which could be implicated in a potential fall.
Root causeMixed-up of materials/components
ActionAll customers who received affected product were initially contacted via phone on April 5, 2012 and made aware of the recall, including instructions to notify customers immediately and arrange replacements. A "Recall Notice" letter dated April 24, 2012 was sent to customers describing the problem and the recommended actions. Contact reference person is Kimberly Light, 1 800-728-7950, info@college-park.com.
Quantity11
DistributionWorldwide Distribution - USA, including the states of CA, MA, ME, WA, TX, VA, AZ and the country of England.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.