Recall Z-1658-2012 ยท LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Becton Dickinson & Co.
Becton Dickinson Phoenix Update Disks (PUD), catalog # 441107/ 448047, component of BD automated microbiology system, and used with BD Gram Positive MIC or MIC/ID panels. Product Usage: The Phoenix Update Disk (PUD) delivers substrate and drug specific information to be used by the BD Phoenix instrument software in determining organism identification and susceptibility results. Information co
| Date | May 24, 2012 (initiated Apr 10, 2012) |
| Status | Terminated |
| Reason | BD identified an increase in vancomycin-resistant strains that may exhibit erroneous results on all PMIC or PMIC/ID panels and could lead to ineffective treatment regimen in affected patients. |
| Root cause | Nonconforming Material/Component |
| Action | Becton Dickinson (BD) sent an Urgent Field Correction Action letter dated April 2012 and an Update letter dated April 18, 2012 to all affected customers via e-mail and UPS overnight delivery. The letter identified the affected products, problem and actions to be taken. The letter provided specific workaround instructions for device users pending software upgrades. A BD Service Engineer will contact affected customers to arrange for the installation of the updated software when it becomes available. The letter included an Acknowledgement Form to be completed. For questions contact Technic |
| Quantity | 3,454,993 total products (U.S.) |
| Distribution | Worldwide Distribution - USA (nationwide) including the states of: AL, AR, AZ, CA, CO, CT ,DC ,FL , GA, IA, IL, IN, KY, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, AND WY. and the countries of Canada, Australia, Philippines, Thailand, Malaysia, Ind |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K030677 | BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-VANCOMYCIN GRAM POSITIVE Becton, Dickinson & CO, Apr 1, 2003 |