Recall Z-1687-2012 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA, Inc.

e.cam Emission Computed Tomography Systems. Siemens e.cam Emission Imaging Computed Tomography Systems is used for all common nuclear medicine procedures.

DateMay 30, 2012 (initiated May 15, 2012)
StatusTerminated
ReasonSiemens has identified a potential patient safety issue for e.cam and Symbia E camera systems that have had the radial drive motor assembly replaced during a Service action. There is a remote chance that during the motor replacement the radial drive gear may have been replaced incorrectly. If the gear is installed incorrectly the mechanical brake may not engage creating a patient hazard if the sys
Root causeEmployee error
ActionOn May 15, 2012, Siemens Medical Solutions USA, Inc. sent an "Urgent Field Correction Recall" letter dated April 11, 2012 to all affected customers. The letters informed customers of the potential patient safety issue for e.cam and Symbia E camera systems that have had the radial drive motor assembly replaced during a Service action. Customers were informed that a Siemens Service Engineer will be inspecting the radial drive gear on their camera system to ensure it is functioning correctly. For questions or assistance they were instructed to contact Siemens Medical Solutions USA, Inc. at 800-
Quantity137 units
DistributionWorldwide Distribution, US (Nationwide) and the countries of Australia, Belgium, Canada, China, France, Germany, Italy, Japan, Republic of Korea, Mexico, Netherlands, Portugal, Spain, Taiwan and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K952109IP500 NUCLEAR GAMMA CAMERA SYSTEM
Siemens Medical Solutions USA, Inc., Aug 23, 1995