Recall Z-1613-2012 ยท GEH Unit, Cryosurgical, Accessories

Galil Medical, Inc.

Galil Medical, 17G Cryoablation Needle Holder FPRPR 4014, Rx only, Sterile EO. Product Usage: Galil Medicals 17G Cryoablation Needle Holder (Needle Holder) is intended for use only with a Galil Medical Needle Testing Device (NTD). The NTD, a component used when performing cryogenic destruction of tissue with a Galil Medical Cryoablation System, is intended for performing a Needle Integrity and Fu

DateMay 16, 2012 (initiated Apr 12, 2012)
StatusTerminated
ReasonGalil Medical is recalling 17G Cryoablation Needle Holder, FPRPR 4014. This voluntary recall has been initiated due to the possibility of excessive bioburden on the Needle Holders.
Root causeMaterial/Component Contamination
ActionGalil Medical sent an "Urgent Medical Device Recall" letter dated April 12, 2012 to all affected customers. The letter identified the affected product, problem, actions to be taken. The letter instructs customers to return all unused affected products, use a revised needle preparation procedure and forward and communicate this Recall Notice and instructions. Customers were instructed to complete and return the attached Response Form with returned product or fax to 1+877-510-7757. For any questions contact Galil Medical Customer Service.
Quantity26 kits
DistributionWorldwide Distribution - USA (nationwide) including the states of CA, NY, NC, WA and the countries of CANADA, DENMARK, FRANCE, NETHERLANDS, TAIWAN and UNITED KINGDOM.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060144SEEDNET/SEEDNETGOLD, CRYOTHERA AND CRYO-HIT SYSTEMS
Galil Medical , Ltd., Feb 13, 2006