Recall Z-1523-2012 ยท DTZ Oxygenator, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corp

Terumo Custom Cardiovascular Procedure Kit (manufactured to the design and specifications of the purchaser/user) for the application it is intended by the user. X-COATED CHANGE OUT PACK WITH FX15E Product Code: 74257

DateMay 11, 2012 (initiated Apr 12, 2012)
StatusTerminated
ReasonLoose connections (i.e., connections between PVC circuit tubing and Oxygenator ports) may result in leaks and patient blood loss
Root causeNonconforming Material/Component
ActionTerumo Cardiovascular Systems issued an Urgent Medical Device Recall notification on 4/12/12 to all user facilities/consignees who have received affected devices. The letter identified the affected product, stated the reason for the recall, the potential hazard involved, and the correction. Users are to review the letter, ensure that all users are made aware of the issue, discontinue use of the affected product, and return it to Terumo CVS. In addition, users are to return the response form to the manufacturer. Affected units will be replaced with new units. Contact Terumo CVS Customer Serv
Quantity3 units
DistributionWorldwide Distribution -- US, including the states of AL, AZ, CA, MO, and NY and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051997CAPIOX RX15 HOLLOW, FIBER OXYGENATOR WITH/WITHOUT HARDSHELL RESERVOIR
Terumo Cardiovascular Systems Corp., Sep 15, 2005
K062381CAPIOX RX OXYGENATOR WITH OR WITHOUT RESERVOIR
Terumo Cardiovascular Systems Corp., Aug 24, 2006