Recall Z-1787-2012 ยท CBK Ventilator, Continuous, Facility Use

GE Healthcare, LLC

GE Healthcare, Aestiva/5 7900 SmartVent, anesthesia machine. Used in Datex-Ohmeda Aestiva/5 Anesthesia Systems.

DateJun 17, 2012 (initiated Apr 10, 2012)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential safety issue due to the interlock slide mechanism missing from the 3-vaporizer manifold of your Aestiva 7900 Anesthesia Device. This Potential for two (2) vaporizers delivering agent at the same time. This could manifest as over delivery of a single agent if both vaporizers contain the same agent or as delivery of more than one agent. This m
Root causeProcess control
ActionGE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 10, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Contact GE Healthcare Customer or Technical Support at 800-345-2700 for questions regarding this notification.
Quantity8
DistributionNationwide Distribution - including the states of AL, FL, GA, MO, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023366DATEX-OHMEDA 7900 VENTILATOR ENHANCEMENTS
Datex-Ohmeda, Inc., Oct 7, 2003