Recall Z-1446-2012 ยท DJR Enzyme Immunoassay, Methadone

Alere San Diego dba Biosite Innovacon Hemosense Inc

Triage TOX Drug Screen (MTD) Panel, Model # 94400. Type of Packaging: 25 individually pouched devices in a labeled kit box. The Triage TOX Drug Screen is a fluorescence immunoassay intended to be used with the Triage Meters for the point-of-care qualitative determination of the presence of drug and/or the major metabolites above the threshold concentrations of up to 10 distinct drug classes, incl

DateApr 20, 2012 (initiated Mar 12, 2010)
StatusTerminated
ReasonThe recall was initiated because Biosite Inc. received an increase in customer calls regarding unexpected positive THC results when using the Triage TOX Drug Screen.
Root causeNonconforming Material/Component
ActionBiosite sent a "PRODUCT ADVISORY NOTICE" letter dated March 12, 2010 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter states that customers may continue to use the product. Customers were instructed to complete and fax an enclosed verification form to Technical Services at 1-858-695-7100. Customers with any questions about the information contained in the notification were instructed to contact Biosite Inc. at (888) 246-7483.
Quantity1,258,002 kits
DistributionWorldwide Distribution-USA (nationwide) and the countries of Canada and European Union.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060791TRIAGE TOX DRUG SCREEN, MODEL 94400
Biosite Incorporated, Jun 22, 2006