Recall Z-1611-2012 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Zimmer Inc.
00-5976-031-23, Nextgen Complete Knee System, Prolong highly crosslinked polyethylene Cruciate Retaining (CR) Micro Articular Surface AC Size Striped Yellow 23 mm Height , Sterile R, Rx ,Zimmer, Inc. Warsaw, IN The NexGen Articular Surface serves as a spacer between the baseplate of the Intended Use. tibia and the metal implant that covers the end of the femur. The micro articular surfaces are
| Date | May 14, 2012 (initiated Feb 24, 2012) |
| Status | Terminated |
| Reason | Zimmer has received 5 complaints where a NexGen CR micro articular surface was used with a standard CR femur, even though the compatibility chart and product labeling indicates that these components are NOT compatible. Investigation identified additional instances where a CR micro articular surface and standard CR femur were sold together on the same purchase order. CR micro articular surfaces are |
| Root cause | Labeling design |
| Action | Zimmer sent an Urgent Correction Notice dated March 5, 2012, to all affected customers via telephone, E-mail and letter notifying users of the appropriate compatibility of micro sizes of NexGen CR and CR-Flex Articular Surfaces and informing them of known cases of incompatible devices being implanted, including associated risks. Customers and surgeons were asked to acknowledge receipt by calling 1-888-912-7349. Questions and concerns or help in notiying accounts about this correction were to be addressed to 1-888-548-8514. For questions regarding this recall call 574-372-4807. |
| Quantity | 83,322, all devices |
| Distribution | Worldwide Distribution - USA (nationwide) and the countries of Poland, Taiwan, Denmark, Korea, Singapore, China, Canada, Australia, Japan, France, Italy, UK, Germany, Spain, Belgium, Portugal, Switzerland, Sweden, India, Thailand, Peru, The Netherlands, Malaysia, and Austria. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K003910 | NEXGEN COMPLETE KNEE SOLUTION CROSS-LINKED POLYETHYLENE CRUCIATE RETAINING (CR) ARTICULAR SURFACE COMPONENTS Zimmer, Inc., Mar 19, 2001 |
| K892800 | MG II POROUS TOTAL KNEE SYSTEM Zimmer, Inc., Aug 3, 1989 |
| K933785 | NEXT GENERATION KNEE SYSTEM Zimmer, Inc., Jan 30, 1995 |
| K951185 | NEXGEN COMPLETE KNEE SOLUTION 9MM ARTICULAR SURFACE Zimmer, Inc., Mar 15, 1996 |