Recall Z-1605-2012 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Zimmer Inc.

00-5976-031-12, Nextgen Complete Knee System, Prolong highly crosslinked polyethylene Cruciate Retaining (CR) Micro Articular Surface AC Size Striped Yellow 12 mm Height , Sterile R, Rx ,Zimmer, Inc. Warsaw, IN The NexGen Articular Surface serves as a spacer between the baseplate of the Intended Use. tibia and the metal implant that covers the end of the femur. The micro articular surfaces are

DateMay 14, 2012 (initiated Feb 24, 2012)
StatusTerminated
ReasonZimmer has received 5 complaints where a NexGen CR micro articular surface was used with a standard CR femur, even though the compatibility chart and product labeling indicates that these components are NOT compatible. Investigation identified additional instances where a CR micro articular surface and standard CR femur were sold together on the same purchase order. CR micro articular surfaces are
Root causeLabeling design
ActionZimmer sent an Urgent Correction Notice dated March 5, 2012, to all affected customers via telephone, E-mail and letter notifying users of the appropriate compatibility of micro sizes of NexGen CR and CR-Flex Articular Surfaces and informing them of known cases of incompatible devices being implanted, including associated risks. Customers and surgeons were asked to acknowledge receipt by calling 1-888-912-7349. Questions and concerns or help in notiying accounts about this correction were to be addressed to 1-888-548-8514. For questions regarding this recall call 574-372-4807.
Quantity83,322 all sizes
DistributionWorldwide Distribution - USA (nationwide) and the countries of Poland, Taiwan, Denmark, Korea, Singapore, China, Canada, Australia, Japan, France, Italy, UK, Germany, Spain, Belgium, Portugal, Switzerland, Sweden, India, Thailand, Peru, The Netherlands, Malaysia, and Austria.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003910NEXGEN COMPLETE KNEE SOLUTION CROSS-LINKED POLYETHYLENE CRUCIATE RETAINING (CR) ARTICULAR SURFACE COMPONENTS
Zimmer, Inc., Mar 19, 2001
K892800MG II POROUS TOTAL KNEE SYSTEM
Zimmer, Inc., Aug 3, 1989
K933785NEXT GENERATION KNEE SYSTEM
Zimmer, Inc., Jan 30, 1995
K951185NEXGEN COMPLETE KNEE SOLUTION 9MM ARTICULAR SURFACE
Zimmer, Inc., Mar 15, 1996