Recall Z-1435-2012 ยท HPZ Implant, Eye Sphere

Howmedica Osteonics Corp dba Stryker Craniomaxillofacial

MEDPOR PLUS SST EZ 22mm Sphere Implant, Howmedica Osteonics Corp, 15 Dart Road, Newnan, GA 30265. MEDPOR PLUS Orbital Implants are intended to restore the volume of an enucleated or eviscerated eye globe.

DateApr 18, 2012 (initiated Mar 20, 2012)
StatusTerminated
ReasonThe MEDPOR PLUS SST EZ 22mm sphere implants are not within specification of 22mm +/- 8, therefore implant is not easily released from the syringe.
Root causeProcess control
ActionThe sole distributor was notified by "Urgent Product Recall" letter on March 20, 2012. "Urgent Medical Device Recall" letters were sent to Stryker Customers on March 26, 2012. The letters described the product issue, hazards, and actions recommended to mitigate the risk. Customer number provided: (800) 962-6558.
Quantity31 units
DistributionUS Distribution, the state of Michigan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021357MEDPOR PLUS ORBITAL VOLUME REPLACEMENT IMPANT
Porex Surgical, Inc., Aug 5, 2002