Recall Z-2203-2012 ยท LLZ System, Image Processing, Radiological

GE Healthcare, LLC

GE Healthcare Advantage Workstations, Advantage Worksation version 4.2 and 4.3 with Volume Viewer 2 (version 6.4-6.8) provided with Advance Vessel Analysis (AVA) and Advantage Workstation 4.4 with Volume Viewer 2 (version 6.8 or before) or Volume Viewer 3 (version 7.2 -7.3) provided with Advanced Vessel Analysis (AVA), Model 2378698-2 Product Usage: Advantage Workstation version 4.2: The Advantag

DateAug 13, 2012 (initiated May 24, 2012)
StatusTerminated
ReasonGE Healthcare is informing you of incorrect measurements when using the Aorta analysis protocol or customized protocols. Until a new software version is installed they are asking you not to use the aorta analysis protocol or the customized protocols for an aneurysm. A second issue was also discovered internally. When loading PET data into the Volume Viewer, if slices are missing in a PET serie
Root causeSoftware design
ActionGE Healthcare sent an "Urgent Medical Device Correction" letter dated May 23, 2012 to consignees. The letter was addressed to Hospital Administrators / Risk Managers, Radiology Department Managers & Radiologists. The letter described the affected product details, safety issue, safety instructions and product correction. For more information on this topic, contact a GE Application Specialist or your GE Field Service representative. For questions or concerns regarding this notification, please call one of the following phone numbers: United States: 800 437 1171 Japan: 0120 - 055 - 919 For
Quantity1055
DistributionWorldwide Distribution - US (nationwide) including the states of : AL, AZ, AR, CA, CO, DE, DC, FL, GA, GU, HI, IK, IL, IN, IA, KY, LA, MD, MA, MI, MN, MS, MO,MT, NE,NH,NJ, NM, NY, NC, ND, OH, OK, PA, PR, SC, SD, TN, TX, VA, WA, WV, WI and the countries of YUGOSLAVIA, VIETNAM, VENEZUELA, UNITED KING
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020483ADVANTAGE WORKSTATION 4.1
GE Medical Systems, Feb 28, 2002
K052995ADVANTAGE WORKSTATION 4.3.
GE Healthcare, Nov 8, 2005