Recall Z-1489-2012 ยท ITX Transducer, Ultrasonic, Diagnostic

Pentax Medical Company

EG-3870UTK, Pentax Ultrasound Video Gastroscope. The EG-3870UTK, Ultrasound Video Gastroscope, is intended to provide optical visualization via a video monitor of, ultrasonic visualization of, and therapeutic access to the Upper Gastrointestinal Tract, including, but restricted to the organs, tissues, and sub-systems; Esophagus, Stomach, Duodenum, Small Bowel, and underlying areas. The instrument

DateMay 3, 2012 (initiated Mar 14, 2012)
StatusTerminated
ReasonEndoscope ultrasound balloon evacuation tube may become clogged with patient body fluid if users fail to properly follow instructions for pre-procedure check, operating the suction valve during procedure and pre-cleaning procedure.
Root causeEquipment maintenance
ActionPentax sent an "URGENT MEDICAL DEVICE RECALL" letter on March 14, 2012 to all affected customers. The letter identifies the product, problem, and the actions to be taken by the customers. A response form was included for customers to complete and return to the firm. Contact the firm at 800-431-5880 ext. 2064 for questions regarding this notice.
Quantity194 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090197EG-3870UTK
Pentax Medical Company, May 19, 2009