Recall Z-0125-2013 ยท LLZ System, Image Processing, Radiological

GE Healthcare, LLC

GE Healthcare, Seno Advantage 2.2 workstations with software version 22_01 and 22_02 A medical image review station that allows easy selection, processing, filming and media interchange of multi-modality images from a variety of diagnosis imaging systems.

DateOct 24, 2012 (initiated May 24, 2012)
StatusTerminated
ReasonGE Healthcare has recently become aware of inaccurate Scale factor annotation provided by SenoAdvantage 2.2 workstations that may impact patient safety. This issue may result in a potential miscalculation of size.
Root causeSoftware design
ActionConsignees were sent on 5/24/12 a GE Healthcare "Urgent Medical Device Correction" letter dated May 23, 2012. The letter was addressed to Hospital Administrators / Risk Managers, Radiology Department Managers & Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, and Product Correction. Customers should verify if their workstation is affected. GE Healthcare will correct all affected workstations by providing a software upgrade and Operator Manual update. A GE Healthcare service representative will contact you to arrange this corr
Quantity12
DistributionWorldwide Distribution -- USA (nationwide) including states of: CA, CT, NY, TX, and MS and countries of TAIWAN, ITALY, CANADA, and BAHRAIN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033400SENO ADVANTAGE
Ge Medical Systems W-400, Dec 4, 2003