Recall Z-1498-2012 ยท JWY Manual Antimicrobial Susceptibility Test Systems

Siemens Healthcare Diagnostics, Inc.

MicroScan Dried Overnight Gram Negative Panels; Panel type: Neg Breakpoint Combo Type 34, catalog number: B1017-404. Siemens Healthcare Diagnostics, West Sacramento, CA.

DateMay 8, 2012 (initiated Jun 16, 2011)
StatusTerminated
ReasonThree lots were confirmed to not meet performance claims due to random non-biological contamination which is only apparent (precipated) after inoculation of the panels.
Root causeProcess control
ActionCustomer Notification letters were sent to all customers of the three MicroScan panels by traceable mail. The letter identified the affected product and asked customers to discard any remaining inventory in addition to contacting customer service for replacement product. The letter also recommended that customers consider the need to review previous test results, conduct patient follow-up, and or repeat tests as needed. Customers were to complete and return the customer reply card to confirm that they received and understood the recall communication.
Quantity680 boxes (20 panels per box)
DistributionWorldwide Distribution -- US, Turkey, Poland, and Great Britain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K862140MICRODILUTION PANELS
American Micro Scan, Jul 16, 1986