Recall Z-1960-2012 ยท LLZ System, Image Processing, Radiological

GE Healthcare, LLC

GE Advantage CT/PET Fusion for Advantage Windows Workstation with Volume View CT Colonography/Navigator 2 software. (Version 4.0, 4.1,or 4.2) The Advantage CT/PET Fusion clinical application provides easy comparison of three dimensional (3D) anatomical images from Computed Tomography (CT) and functional images from Positron Emission Tomography (PET).

DateJul 9, 2012 (initiated May 24, 2012)
StatusTerminated
ReasonGE Healthcare has become aware of a potential safety issue due to incorrect alignment of fused CT and PET images associated with the CT/PET fusion capabilities of your Volume Viewer and CT/PET Fusion application. Volume Viewer allows two studies (one PET and one CT) that are registered in the same spatial domain to be viewed simultaneously in fused mode. The two images should be aligned to th
Root causeSoftware design
ActionGE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated May 23, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter provides safety instructions for customers to use until a software upgrade is available. Contact the firm at 800-437-1171 for questions regarding this recall.
Quantity70
DistributionWorldwide Distribution-USA (nationwide) including the states of CA, FL, IL, MA, NJ, NY, SC, and TX and the countries of TURKEY, TUNISIA, TAIWAN, SPAIN, SINGAPORE, SAUDI ARABIA, RUSSIAN FEDERATION, KENYA, JAPAN, ITALY, IRELAND, INDIA, GREAT BRITAIN, GERMANY, FRANCE, EGYPT, CHILE, CHINA, CANADA, BRAZI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010336ADVANTAGE WINDOWS CT/PET FUSION
Ge Medical Systems, Inc., Feb 16, 2001
K012313CT COLONOGRAPHY/NAVIGATOR2
Ge Medical Systems, Inc., Aug 7, 2001