Recall Z-1486-2012 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

Siemens-branded linear accelerator with a 550 TxT Treatment Table with a serial number <_ 1894. Siemens brand ARTISTE, ONCOR and PRIMUS Linear Accelerator systems include the 550 TxT" Treatment Table. Manufactured by: SIEMENS AG, MEDICAL SOLUTIONS Roentgenstrasse 19-21 Kemnath, GERMANY 95478. Product Usage: The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linear

DateMay 3, 2012 (initiated Apr 3, 2012)
StatusTerminated
ReasonSIEMENS has received reports where a dropping down of the 550 TxT" Treatment Table was observed while the table was being moved downwards during operation by means of the hand control.
Root causeSoftware design
ActionSiemens sent Customer Information letters to all affected customers beginning April 3, 2012 via Siemens Field Service Engineers. The letter identified the affected product, what is the issue and what has Siemens done to address this issue. The letter included an Update Instructions TH006/12/S - TxT Table Software Update VD00A. SIEMENS has arranged for a modification of the affected system. A SIEMENS Service Representative will ensure that the new software version VD00A is installed on affected system. The letter instructs customer to please include this Customer Information in their Digita
Quantity624 active devices
DistributionWorldwide Distribution - USA (nationwide) including P.R and the countries of: Africa, Angola, Argentina, Australia, Austria, Belgium, Bosnia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Egypt, France, Germany, Greece, Hungary, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malays
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060226ONCOR EXPRESSION
Siemens Medical Solutions USA, Inc., Mar 15, 2006
K103606ARTISTE SOLUTION WITH SYS_VB50 UPDATE
Siemens Medical Solutions USA, Inc., Apr 15, 2011