ClearanceBrief
Recall Z-1536-2012 ยท LLI Bed, Therapeutic, Ac-Powered, Adjustable Home-Use

GF Health Products, Inc./d.b.a.Basic American Metal Products

LUMEX A Graham-Field Brand, Bed End Sert. Low/Full-electric, Plastic Panels, Item 690-7604-000 Manufactured for : GF Heath Products, Inc., Atlanta, Georgia 30360 Made in China. LUMEX A Graham-Field Brand, Bed End Sert. Low/Full-electric, Walnut Panels, Item 690-7004-000 Manufactured for : GF Heath Products, Inc., Atlanta, Georgia 30360 Made in China. Therapeutic bed intended for home use.

DateMay 11, 2012 (initiated Apr 17, 2012)
StatusTerminated
ReasonGF Health Products, Inc., is conducting a recall on the Lumex Patriot Full Electric Homecare "Bed Ends" manufactured 3/27/09, 5/9/09, 5/30/09 and 6/20/09. The supplier deviated from the print specifications of a machined pin to a sheared pin. This deviation has the potential to provide room for the pin to move during use. If failure were to occur, the bed could descend rapidly causing non-typic
Root causeComponent change control
ActionGF Health Products, Inc. sent an "URGENT CORRECTION AND REMOVAL" letter dated April 17, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to immediately cease use of the affected product, secure the product, and to prevent the product from further distribution to customers and end users. Contact the firm at 1-800-556-6508 for questions regarding this notice.
Quantity3095 pairs of bed ends
DistributionWorldwide Distribution-USA (nationwide) including the states of AL, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, and WI and the countries of EL SALVADOR, CHILE, GUATEMALA, TRINIDAD, TOB
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.