Recall Z-1546-2012 ยท DHR System, Test, Rheumatoid Factor

Thera Test Laboratories, Inc.

RF IgA Calibrator, a component of the EL-RF/3 Rheumatoid Factor Test Kit; an in vitro diagnostic; TheraTest Laboratories, Lombard, IL 60148; component catalog #12833, kit catalog #303-305. The TheraTest EL-RF/3 is intended as an aid to the diagnosis of Rheumatoid Arthritis (RA).

DateMay 14, 2012 (initiated Mar 29, 2012)
StatusTerminated
ReasonThe RF IgA Calibrator component of the EL-RF/3 Rheumatoid Factor test system may be unstable, resulting in a drop in the RF IgA calibrator OD value which could invalidate an RF IgA test run.
Root causeNonconforming Material/Component
ActionTheraTest sent "Urgent Device Recall" letters dated March 28, 2012 to customers that received the affected kit, requesting them to return or destroy the remainder of the product. TheraTest issued a replacement kit to all customers that had unused product as well as any material they consumed which resulted in a failed run. Customers were also instructed to complete the enclosed confirmation form and fax the completed form back to TheraTest. Any questions were directed to TheraTest at 800-441-0771.
Quantity207 kits
DistributionNationwide Distribution (US)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K941398EL-RF/3 (IGM-IGG-IGA) KIT
Theratest Laboratories, Inc., Jul 28, 1995