Recall Z-1383-2012 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Siemens Medical Solutions USA, Inc

Siemens Axiom Luminos dRF solid state x-ray imager Solid state x-ray imager (flat panel/dig. imager)

DateApr 9, 2012 (initiated Feb 16, 2012)
StatusTerminated
ReasonDuring regular product monitoring, Siemens became aware of a potential issue with the AXIOM Luminos dRF. Firm became aware of a potential issue with torque wrenches - during a regular scheduled calibration of the wrench, it was discovered that the wrench was out of the specification. The wrench was over-torquing approximately 21% for low-end readings (50 ftlbs) and 9% for high-end readings (250
Root causeNonconforming Material/Component
ActionSiemens initiated a recall/correction on February 16, 2012 by issuing Update Instruction XP056/11/S to the affected customers. All screws on the unit support on AXIOM Luminos dRF units will be replaced by a Siemens service engineer. For information regarding this recall call 610-219-4834.
Quantity2
DistributionNationwide Distribution including California
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062623AXIOM LUMINOS DRF
Siemens Medical Solutions USA, Inc., Aug 22, 2007