Recall Z-1365-2012 ยท MUJ System, Planning, Radiation Therapy Treatment

Computerized Medical Systems Inc

Monaco Radiation Treatment Planning System, Version 3.10.00. Planning of radiation therapy.

DateApr 3, 2012 (initiated Mar 5, 2012)
StatusTerminated
ReasonAn unnecessary dose refresh can occur in certain situations which causes inconsistencies in the Beam Doses and the Total Dose.
Root causeSoftware design
ActionAn "IMPORTANT SAFETY NOTICE" was issued between 3/5-9/12 via e-mail, FAX, or traditional mail. The Notice described the correct functioning of the system, the problem, when and why the issue occurs, the clinical impact, and the workaround. Patch 3.10.01 is now available. Customers can download their patch at the link provided in the Notice. The Notice also provided a list of offices where customers can contact if they have questions or further support is needed.
Quantity44
DistributionWorldwide Distribution -- USA, including the states of OH, IN, WI, CA, FL, IL, LA, NJ, MA, NC, and MO and the countries of Canada, Australia, Belgium, China, France, Germany, Japan, The Netherlands, New Zealand, Portugal, Switzerland, Taiwan, and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110730MONACO RTP SYSTEM
Computerized Medical Systems, Inc., Jun 24, 2011