Recall Z-1501-2012 ยท GHM Test, Sickle Cell
MichClone Associates, Inc.
Sickle Heme, Hemoglobin S Solubility For Qualitative Determination of Hemoglobin S Test; ,100 Test Kit Product, No MCA 100, Contents: Two bottles of Sickle Cell Buffer, Two Bottles of Sickle Cell Powder; One Line Scale Two Dispensing Closures; For Invitor Diagnostics Use, Store at Room Temperature, Michclone Associates, Madison Heights, MI 48071. For the qualitative determination of positive or n
| Date | May 9, 2012 (initiated Feb 23, 2012) |
| Status | Terminated |
| Reason | Michclone Associate's has received two complaints concerning the powder vial included in the 100 test kit observed producing pressure while sealed. When the powders are opened, the pressure from within the vial releases a small amount of powder into the air. This powder. if released near the face and inhaled can cause minor irritation resulting in sneezing and nasal congestion. Investigation f |
| Root cause | Nonconforming Material/Component |
| Action | Michclone Associates, Inc. sent an "URGENT MEDICAL DEVICE RECALL" letter dated March 15, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to identify affected lots, notify sub accounts, and complete and return the response form to arrange return of unused product and replacement . Forms can be returned by FAX 248(307)-9498, mail: Michclone Associates Inc. 680 Ajax Drive, Madison Heights, MI 48118 or Email: toben@greatlakesdiag.com. Questions and concerns should be addressed to Technical Su |
| Quantity | 120 kits |
| Distribution | Nationwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K860812 | SICKLE CELL REAGENT SET Michclone Associates, Inc., Apr 11, 1986 |