Recall Z-1283-2012 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Philips Medical Systems Gmbh, DMC

Philips Easy Diagnost Eleva DRF, Software release 4.0 Fluoroscopic image-intensified Stationary X-Ray System.

DateMar 23, 2012 (initiated Mar 9, 2012)
StatusTerminated
ReasonUnderexposed image due to lag in switching between auxiliaries
Root causeSoftware design
ActionPhilips Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated February 27, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The firm provided a work around for customers to use until a software update can be installed. Customers should contact their local Philips representative for questions regarding this notice.
Quantity88 units
DistributionWorldwide Distribution-USA (nationwide) and the countries of Australia, Germany, Indonesia, Malaysia, Saudi-Arabia, Singapore, and Thailand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031535PHILIPS EASYDIAGNOST ELEVA
Philips Medical Systems North America Co., Jun 17, 2003