Recall Z-1385-2012 ยท KXK Source, Brachytherapy, Radionuclide

Varian Medical Systems, Inc.

VariSeed 7.1, software application, part number H62, manufactured by Varian Medical Systems. VariSeed 7.1 is a computer based software application for planning and evaluating prostate brachytherapy procedures.

DateApr 10, 2012 (initiated Jan 25, 2012)
StatusTerminated
ReasonBrachytherapy device requires the user to enter radioactive source activity, but software may specify unintended units for activity and lead to overdose in patient treatment.
Root causeLabeling design
ActionVarian Medical Systems notified direct account healthcare practitioners/ medical physicists with "Urgent Medical Device Correction/Urgent Field Safety Notice" letters by mail on 01/19/2012. Varian's advisory described how to mitigate user error and the firm's intentions to revise Instructions for Use. Technical Support phone numbers were provided for US and international customers.
Quantity1377 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov