Recall Z-1250-2012 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

Medtronic Inc. Cardiac Rhythm Disease Management

Medtronic EnTrust, Dual Chamber Implantable, Cardioverter Defibrillator with Ventricular Therapies, models D153DRG and D154DRG. The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment for automated treatment of life-threatening ventricular arrhythmias. Sterile. Medtronic, Inc., Minneapolis, MN 554

DateMar 22, 2012 (initiated Mar 6, 2012)
StatusTerminated
ReasonA small percentage of EnTrust ICDs may not meet expected longevity or provide at least three months of device operation between the Elective Replacement Indicator (ERI) and End of Life (EOL) due to a more-rapid-than-expected drop in battery voltage.
Root causeDevice Design
ActionAn "Urgent Medical Device Correction" letter was sent to physicians beginning 3/6/2012. The letter described the issue and provided patient management recommendations. Additionally the letter stated that ongoing updates to the product performance report will be posted on a Medtronic website.
Quantity68,804 devices worldwide
DistributionWorldwide Distribution -- USA, and the countries of: Argentina, Aruba, Australia, Austria, Bahamas, Barbados, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India,