Recall Z-1362-2012 ยท CBK Ventilator, Continuous, Facility Use

Maquet Inc.

Servo-i ventilator system Product Usage: Intended for treatment and monitoring of patients in the range of neonates, infants and adults with respiratory failure or respiratory insufficiency.

DateApr 3, 2012 (initiated Dec 22, 2011)
StatusTerminated
ReasonThe Servo-i and Servo-s ventilator systems equipped with specific revision os of the AC/DC convertor can result in electric shock during servicing.
Root causeDevice Design
ActionMaquet sent a Product Recall Urgent-Medical Device Corrective Action letter dated December 22, 2011 to all affected customers. The letter identified the affected products, patient risk, product issue and actions to be taken. Customers were instructed to complete and return the attached Customer Fax Response form as soon as possible. Maquet Field Service Technicians will update all affected products at all the concerned customer sites. Customers are advised untill Maquet has updated all the affected products (all of which are still under warranty), inform all the concerned biomedical staff
Quantity647 US; 2462 Foreign; 27 Spare Parts Foreign
DistributionUSA Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041223MODIFICATION TO SERVO-I VENTILATOR SYSTEM, MODEL 64 87 800 E407E
Maquet Critical Care AB, Jul 29, 2004