Recall Z-1245-2012 ยท LFL Instrument, Ultrasonic Surgical
Ethicon Endo-Surgery Inc
ENSEAL Laparoscopic 5 mm Diameter Tissue Sealer G2 35 cm Length Curved Jaw, Model # NSLG2C35 and ENSEAL Laparoscopic 5 mm Diameter Tissue Sealer G2 45 cm Length Curved Jaw, Model # NSLG2C45. Ethicon Endo-Surgery, LLC Guaynabo, Puerto Rico 00969 The ENSEAL G2 Tissue Sealers are indicated for bipolar coagulation and mechanical transection of tissue during laparoscopic and open procedures. The devi
| Date | Mar 21, 2012 (initiated Jan 26, 2012) |
| Status | Terminated |
| Reason | Ethicon Endo-Surgery initiated a voluntary global recall for specific production lots of ENSEAL G2 Curved and Straight Tissue Sealers due to two potential issues that may occur related to the activation button: (i) Continuous Activation: The ENSEAL device/system may continue to activate when the activation button is released prior to reaching the end of cycle (tone 3). (ii) No Activation |
| Root cause | Nonconforming Material/Component |
| Action | Ethicon Endo-Surgery sent an Urgent Medical Device Recall letter dated February 10, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately stop use of the affected product, fill out the attached Business Reply Form (BRF) and return to and return the affected product to: Ethicon Endo-Surgery ATT: ENSEAL G2 Recall 4545 Greek Road Cincinnati, Ohio 45242 Customers were instructed to choose one of the following response options: Return the Business Reply Form to their sales represen |
| Quantity | 633 pieces |
| Distribution | Worldwide Distribution - USA including AZ, CA, CT, FL, IL, IN, KS, KY, MA, MI, MO, NC, NY, OH, PA, TN & TX and the countries of Arab Emirates, Austria, Belgium, France, Germany, Great Britain, Italy, Jordan, Kingdom of Saudi Arabia, Lebanon, Portugal, Slovenia, Sweden and Turkey. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K112033 | ENSEAL G2 TISSUS SEALERS Ethicon Endo-Sugery, LLC, Nov 8, 2011 |