Recall Z-1370-2012 ยท LLZ System, Image Processing, Radiological

Hologic, Inc.

Hologics SecurView DX workstation, software version 7.3.0 intended for viewing mammography images and other imaging modalities. SecurView DX is intended for selection, display, manipulation, filing and media interchange of multimodality images from a variety of different modality systems. It also interfaces to various image storage and printing devices, using DICOM or similar interface standards

DateApr 5, 2012 (initiated Feb 28, 2012)
StatusTerminated
ReasonSecurView DX with software version 7.3.0 defects when used to view mammography tomosynthesis images
Root causeSoftware design
ActionHologic Inc. sent an "URGENT: MEDICAL DEVICE RECALL" letter dated February 28, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter provides instructions on how to re-review the stored images and correct any annotation defect. A SecurView DX 7.3.0 Confirmation of Notification form was enclosed for customers to complete and return to the firm via fax at 866-652-8674. Contact Hologic's Help Desk at 1-877-371-4372 for questions regarding this notice.
Quantity7 units
DistributionWorldwide Distribution-Distributed in the state of Texas and the countries Belgium, Brazil, and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103385SECURVIEW DX DIAGNOSTIC WORKSTATION
Hologic, Inc., Feb 2, 2011