Recall Z-1204-2012 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Stryker Sustainability Solutions, Inc.

The Reprocessed Mitek Arthroscopic Wand is considered a soft tissue ablators. Soft tissue ablators are radiofrequency (RF) surgical tools designed for removal and dissection of tissue in arthroscopic surgeries. A generator or controller serves as the power unit. Soft tissue ablators are available in a wide range of sizes and angles; some models include thermal or suction functions. The outer box

DateMar 13, 2012 (initiated Jun 1, 2009)
StatusTerminated
ReasonOn 06/09/2009, Stryker Sustain ability Solutions, formerly Ascent Healthcare Solutions, recalled lots 674201 & 674290 of their Reprocessed Mitek Arthroscopic Wand due to the lots not being tested in accordance with the functional test requirement during the time frame of 04/14/09 to 05/25/09.
Root causeProcess control
ActionAscent Healthcare Solutions notified their customers by presenting a letter dated June 1, 2009 from an Ascent Representative. The representative was to recover any products that may have been affected. Questions were directed to 1-888-888-3433.
Quantity48
DistributionNationwide Distribution (USA), including the states of: TX, NY, MN, CA, and CT.