Recall Z-1358-2012 ยท EHD Unit, X-Ray, Extraoral With Timer

Gendex Dental Systems

Gendex eXpert DC Intraoral Dental X-Ray System Device is to be used as an extra-oral source of X-Rays in Dental Radiography

DateApr 2, 2012 (initiated Dec 29, 2011)
StatusTerminated
ReasonImproper hardware was used during installation of the eXpert DC Intraoral X-Ray system units causing drifting of the tubehead assembly.
Root causeNonconforming Material/Component
ActionGendex Dental Systems sent a Urgent Medical Device Correction letter dated December 29, 2011, to both affected customers. The Customers were provided background information and details on how the nonconforming installations of the devices would be corrected. Gendex Dental Systems apologizes for any inconvenience caused by this issue, and request your corporation as we correct this problem. The customers were instructed to complete and return the Acknowledgement Form upon receipt of the recall notification letter and fax to (215) 997-5665. For further questions please call (215) 954-0365
Quantity52
DistributionUSA including the states of Iowa and Illinois.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K992610GENDEX 765DC, MODEL 110-0154
Dentsply Intl., Sep 7, 1999