Recall Z-1331-2012 ยท LEA Preservative, Cytological

Leica Biosystems Richmond Inc.

Surgipath Cyto Jar Fixative; a 120 mL clear, leakproof container prefilled with 45 mL of preservative containing polyethylene glycol for the convenient collection, fixation and preservation of all cytologic specimens, 50 jars per case; for in vitro diagnostic use only; Leica Biosystems, 5205 Rt. 12, Richmond, IL 60071; Made in USA The Cyto Jar is for the convenient collection, fixation and preser

DateMar 29, 2012 (initiated Feb 20, 2012)
StatusTerminated
ReasonProduct was distributed with the incorrect lot number and expiration date. The lot number and expiration date should have been listed at 073013 and instead was noted as 073014 which is beyond the maximum expiration dating of 18 months for this product.
Root causeProcess design
ActionThe firm, Leica Microsystems, Inc., sent an "URGENT MEDICAL DEVICE RECALL" letter dated February 17, 2012, via FedEx on February 20, 2012 to all affected customers. The letter describes the product, problem, and actions to be taken. A second letter was sent to the customers on February 29, 2012, clarifying that the incorrect lot number and expiration date was 073014, and should have been 073013. The customers were instructed to discontinue use of the product; return the product and complete and return the enclosed Acknowledgement Form via fax to 815-678-2216 within one week of receiving the
Quantity11 cases
DistributionNationwide distribution: USA including states of: Michigan, Missouri and New York.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.