ClearanceBrief
Recall Z-1222-2012 ยท LXH Orthopedic Manual Surgical Instrument

Biomet U.K., Ltd.

Vanguard Knee Instrumentation Distal Cutting Block without Handles, REF 32-487002 LOT 2561329, Biomet UK Ltd, South Wales, UK Used to guide a saw blade in resecting the the distal femor. Intended to be used with a femoral alignment guide guide.

DateMar 19, 2012 (initiated Feb 14, 2012)
StatusTerminated
ReasonThis recall was initiated following a report that a bag labeled as REF. 32-487557/Lot no 2561334 (Vanguard Tibial Resection head universal w/ alignment tower) actually contained a Ref. 32-487002/Lot no. 2561329 (Vanguard Distal cutting Block w/out handle attachment). Investigation found packaging of the affected product may contain the incorrect item.
Root causePackaging change control
ActionBiomet U.K. sent a Field Safety Corrective Action notice dated February 14, 2012 to all affected customers. The notice identified the affected products, problem and actions to be taken. Customers were instructed to locate, discontinue use and quarantine the recalled products pending return to Biomet. The letter recommended customers provide the notice to any other organization who may have received the affected instruments. Customers were instructed to complete and return without delay the attached "Fax-Back Response Form" acknowledging receipt of notice and indicating the quantity of prod
Quantity4 units
DistributionWorldwide Distribution - in the countries of France, Japan, Italy, Netherlands, UK , no devices distributed in the USA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.