Recall Z-1387-2012 ยท MMI Immunoassay Method, Troponin Subunit
Roche Diagnostics Operations, Inc.
Elecsys Troponin I Immunoassay For the in-vitro quantitative determination of cardiac troponin I in human serum and plasma. Elecsys Troponin I is intended to aid in the diagnosis of myocardial infarction.
| Date | Apr 11, 2012 (initiated Mar 12, 2012) |
| Status | Terminated |
| Reason | An investigation has confirmed a considerably reduced recovery of Li-heparin plasma samples for Troponin I lot 163176 and Troponin I STAT lot 163177. In the worst case, Li-Heparin sample recovery may be as low as 50% of the serum recovery. Correct serum recovery has been confirmed for the affected lots. The issue is caused by the raw material lot (Poly-L-Lysin) instability. Poly-L-Lysin is the |
| Root cause | Nonconforming Material/Component |
| Action | The firm, Roche Diagnostics Corporation, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated March 12, 2012 to all customers that received the affected lots of Elecsys Troponin I or Elecsys Troponin I STAT. The letter describes the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of the affected product; discard the affected product from their inventory according to their site's local regulations; if their facility has distributed the affected product to other sites, ensure that this letter is provided to those sites; complete and return t |
| Quantity | 739 kits |
| Distribution | Worldwide distribution. Argentina, Kenya, Austria, South Korea, Belgium, Kuwait, Bahrain, Morocco, Brazil, Mexico, Switzerland, Nigeria, Chile, New Zealand, Columbia, Philippines, Germany, Pakistan, Algeria, Poland, Egypt, Qatar, Spain, Tunisia, France, Turkey, United Kingdom, Taiwan, India, Ukrai |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K082699 | ELECSYS TROPONIN I AND TROPONIN I STAT TEST SYSTEMS Roche Diagnostics, Aug 19, 2009 |