Recall Z-1227-2012 ยท GDW Staple, Implantable

Covidien LP

Covidien Duet TRS Articulating 60- 3.5MM, Single Use Loading Unit Ref: DUET6035A The ENDO GIA Universal Staplers with DUET TR S Loading Units have applications in abdominal, gynecologic and pediatric surgery for resection, transection, and creation of anastomosis. They may be used for transaction and resection of liver substance, hepatic vasculature and biliary structures and for transection and

DateMar 19, 2012 (initiated Feb 16, 2012)
StatusTerminated
ReasonMissing component results in the stapler not firing.
Root causeProcess control
ActionThe firm, Covidien, sent a "Medical Device Recall" letter dated February 16, 2012 to customers via federal express. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of Lot# N1H003LX of Duet TRS Articulating Loading Unit and segregate affected product; return inventory to Field Returns Department, 195 McDermott Road, North Haven, CT 06473 and contact customer service at SDFeedback@Covidien.com or (800) 722-8772, option #1 for RGA # prior to returning; and complete and return the Recall Product Return Form via fax
Quantity1158 units
DistributionWorldwide distribution: USA (nationwide) and country of: Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080898AUTO SUTURE ENDO GIA STAPLERS WITH ENDO GIA SINGLE USE LOADING UNITS WITH STAPLE LINE REINFORCEMENT
Tyco Healthcare Group LP, May 5, 2008