Recall Z-1184-2012 ยท IXI Block, Beam-Shaping, Radiation Therapy

Siemens Medical Solutions USA, Inc

Siemens mMLC (ModuLeaf) safety bar and trolley label. Siemens Medical Solutions USA, Inc. Concord, CA. The intended use of the Siemens branded mMLC is an accessory used with the Linear Accelerator systems to deliver x-ray photon and electron radiation for the therapeutic treatment of cancer. This accessory shapes the radiation beam, per the leaf parameters, to minimize the amount of dose to outl

DateMar 7, 2012 (initiated Feb 7, 2012)
StatusTerminated
ReasonThis correction is being performed to update the safety knob of the Siemens brand mMLC (ModuLeaf) to conform to tie IEC 60601-1 safety standard.
Root causeNonconforming Material/Component
ActionThe firm, SIEMENS, issued a "Customer Information" letter to its customers. The letter described the product, problem and actions to be taken. Siemens developed a new steel knob that complies with the latest standards; Siemens checked all ModuLeaf mMLC trolleys and if necessary refitted with the safety label and also arranged for an inspection and modification of the customers system- a Release of the safety update UI TH 026/11/S to distribute and hardware update commenced on February 7, 2007 by certified mail. If you have any questions, contact the Regulatory Specialist II at 925-602-8175.
Quantity46 active devices
DistributionWorldwide distribution: USA (nationwide) including states of: CO, IA, KY, NJ, OH and TX; and countries of: Algeria, Austria, Belgium, Brazil, China, Egypt, France, Germany, Hungary, India, Italy, Japan, Kuwait, Malaysia, Mexico, P.R., Russian Fed., Saudi Arabia, South Africa, Spain, Turkey, and Uni
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030609MODULEAF
Mrc Systems GmbH, Mar 17, 2003