Recall Z-1268-2012 ยท MRG System, Test, Vitamin D

Siemens Healthcare Diagnostics, Inc

Siemens ADVIA Centaur Vitamin D Diluent 2-pack, Reference 10494100 in vitro diagnostic use in the quantitative determination of total 25 (OH) vitamin D in human serum and plasma (EDTA, lithium-heparin, sodium heparin) using the ADVIA Centaur and ADVIA Centaur XP systems The ADVIA Centaur Vitamin D Total (Vit D) assay is for in vitro diagnostic use in the quantitative determination of total 25 (OH

DateMar 21, 2012 (initiated Feb 16, 2012)
StatusTerminated
ReasonAutomatic and system-performed dilution calculation factor for Vitamin D dilutions is incorrect and causes diluted patient samples to under recover by approximately 50%.
Root causeNonconforming Material/Component
ActionThe firm, Siemens Healthcare Diagnostics, sent an "Urgent Device Recall Notice" on February 16, 2012 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to discontinue using the automatic and system-performed dilution features for Vitamin D; to discontinue the use of ADVIA Centaur VitD Diluent, 2-pack (REF 10494100); to manually dilute samples with values greater than the assay range of 150 ng/mL (375 nmoI/L), using the ADVIA Centaur VitD Diluent, 1 bottle (REF 10632114) as stated in the Vitamin D assay Instructions For Use (IFU)
Quantity442 units
DistributionWorldwide distribution: USA (nationwide) and countries of: Columbia, China, Denmark, Egypt, and Hong Kong.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110586ADVIA CENTAUR VITAMIN D TOTAL ASSAY