Recall Z-1239-2012 ยท JJX Single (Specified) Analyte Controls (Assayed And Unassayed)

Beckman Coulter Inc.

Access Ostase QC, Part Number: 37309. The Access Ostase QC is intended for monitoring system performance of the Access Ostase assay.

DateMar 21, 2012 (initiated Nov 10, 2011)
StatusTerminated
ReasonThe recall was initiated because Beckman confirmed that the Access Ostase Calibrator and QC lots do not meet expiration date claims. The Access Ostase Reagent Kit lots identified were quality control tested and released using Access Ostase Calibrator and QC lots that contain the implicated raw material.
Root causeProcess design
ActionBeckman Coulter sent an "URGENT PRODUCT CORRECTION" letter dated November 9, 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue use and discard all remaining inventory. In addition, a Response Form was enclosed for customers to complete and return. Contact Beckman's Customer Support Center at (800) 854-3633 for questions regarding this notice.
Quantity4961 units total (1321 units in US)
DistributionWorldwide Distribution-USA (nationwide) and the countries of Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, France, Georgia, Germany, Greece, Hong Kong, Hungary, India,Ireland, Italy, Japan, Jordan, Lebanon, Liechtenstein, Mexico, Morocco, Netherl