Recall Z-0613-2013 ยท JIS Calibrator, Primary

Beckman Coulter Inc.

Access Immunoassay Systems Total T4 Calibrators, Part Number: 33805. The Access Total T4 Calibrators are intended to calibrate the Access Total T4 assay for the quantitative determination of total thyroxine (T4) levels in human serum and plasma using the Access Immunoassay Systems. Total T4 assay results are used for diagnosis and confirmation of thyroid disorders such as hyper- and hypothyroidi

DateDec 22, 2012 (initiated Jul 27, 2011)
StatusTerminated
ReasonA recall communication was initiated because Beckman has identified that Access Total T4 calibrator lots 019818, 021654, and 024072 do not meet their 12 month expiration date claim.
Root causeNonconforming Material/Component
ActionA recall communication was initiated on July 27, 2011 with Beckman forwarding a Product Corrective Action (PCA) letter with attached PCA response form to all their customers who purchased the Access Immunoassay Systems Total T4 Calibrators. The letter provides the customers with an explanation of the problems identified and actions to be taken. Customers were instructed to complete and return the enclosed response form within 10 days. Customers can call for product replacement in the US, by contacting Client Services at (800) 526-3821, option 1. Customers in Canada can contact Customer ser
Quantity6,634 units total (2,742 units in US)
DistributionWorldwide Distribution-USA (nationwide) and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Egypt, El Salvador, Ethiopia, France, Germany, Greece, India, Indonesia, Ireland, Italy, Japan, Jordan, Kazakhstan, Le
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023369ACCESS TOTAL T4 ASSAY
Beckman Coulter, Inc., Dec 9, 2002