Recall Z-1216-2012 ยท KPR System, X-Ray, Stationary

Philips Healthcare Inc.

Philips EasyDiagnost Eleva DRF, Fluoroscopic x-ray system Software Release 3.0 and 3.1 Model: 706034 Product Usage: Radiological image processing system (computed radiography x-ray system)

DateMar 15, 2012 (initiated Feb 14, 2012)
StatusTerminated
ReasonMisinterpretation of Mirror Icon-Mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system
Root causeSoftware design
ActionPhilips Healthcare sent an Urgent Medical Device Correction letter dated February 8, 2012 to all affected customers. The letter identified the affected product, problem description, hazard involved, actions to be taken by customer/user and actions planned by Philips. The letter contains important information for continue safe and proper use of the equipment. Philips will contact customers to schedule the installation of the updated software. For further information or support contact your local Philips representative at the Technical Support Line 866-767-2822.
Quantity60 units
DistributionWorldwide Distribution - USA Nationwide