ClearanceBrief
Recall Z-1230-2012 ยท DQK Computer, Diagnostic, Programmable

Biosense Webster, Inc.

CARTO 3 Systems, Product Code M-4800-01 The intended use of the CARTO 3 System is catheter-based cardiac electrophysiological (EP) procedures.

DateMar 20, 2012 (initiated Feb 9, 2012)
StatusTerminated
ReasonSome CARTO 3 Systems may get radio frequency interference through certain interconnected cable configurations resulting in jumping or disappearing catheter icon and/or error messages during ablation.
Root causeNonconforming Material/Component
ActionThree Urgent Customer Notification recall communications were initiated 2/9/12 with Biosense forwarding the notification to customers with affected backplane cards (electrophysiology lab managers, a notification for customers with affected back plane cards (risk managers), and a notification for customers with affected backplane cards and smarttouch module along with customer acknowledgement forms. The letters provides the customers with an explanation of the problem identified and steps to take in the event that the issue is observed. If the steps provided do not resolve the issue, customer
Quantity134 units total (32 units in US)
DistributionWorldwide Distribution -- US, Japan, China, Ireland, Germany, Italy, Slovenia, Indonesia, Poland, Spain, United Kingdom, Switzerland, Belgium, and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090017CARTO 3 EP NAVIGATION SYSTEM, V1.0 AND ACCESSORIES
Biosense Webster, Inc., Oct 13, 2009
K093455CARTO 3 VERSION 1.05 EP NAVIGATION SYSTEM AND ACCESSORIES
Biosense Webster, Inc., Dec 4, 2009
K103746CARTO 3 EP NAVIGATION SYSTEM, VERSION 2.0 (W/STANDARD LOCATION PAD),CARTO3 EPNAVIGATION SYSTEM,VERSION 2.0 W/RMT PAD)
Biosense Webster, Inc., Jan 21, 2011
K112007CARTO 3 EP NAVIGATION SYSTEM, VERSION 2.2
Biosense Webster, Inc., Oct 5, 2011